APPUED MICROBIOLOGY, Nov. 1969, p. 861-868
Copyright © 1969 American Society for Microbiology
Vol. 18, No. 5
Printed in U.S.A.
Automated Instrument for the Fluorescent
Treponemal Antibody-Absorption Test
and Other Immunofluorescence Tests
GERALD F. BINNINGS, MEL J. RILEY, MERRITT E. ROBERTS, RICHARD BARNES,
AND THOMAS C. PRINGLE
Medical Instrumentation, Chemical & Biological Department, Space Division, Aerojel General Corporation,
El Monte, California 91734
An automated diagnostic test instrument and its development program are described. The instrument automates the fluorescent treponemal antibody-absorption
test for syphilis to the extent that only 4 hr of technician time is required to conduct
approximately 200 tests daily. Evaluation to date suggests its efficacy. In addition,
preliminary studies indicate the feasibility of detecting antibodies to Toxoplasma
gondii, Plasmodium malariae, and nucleoprotein (antinuclear factor). The instrument would seem to have broad application for routine and research immunofluorescence testing. Two elements comprise the instrument: a slide processor and a microscope attachment. The slide processor is an electro-pneumatically actuated device
which automatically feeds special laboratory slides, on which antigen or other
reagents are prefixed, through a series of operations which provide reagent application, incubation, washing, drying, and stacking of the finished slides for readout.
The instrument provides flexibility in that incubation time and temperature as well as
point, sequence, and duration of reagent application can be varied to accommodate
a variety of immunofluorescence techniques. The microscope attachment can be
fitted to all conventional dark-field fluorescence microscopes and makes possible
the reading of three to six slides per minute. The reacted slides from the processor
are injected sequentially onto the stage of the microscope by movement of a lever.
As injected, slides are automatically in visual focus; fine focus is occasionally required. Scanning of the reacted field is accomplished by means of the normal
microscope controls. A buffered glycerol coupling is maintained between the darkfield condenser substage lens and the slide cover glass by means of a pushbuttonactuated feed system.
Aerojet-General Corp., Chemical & Biological
Division, under contract to Department of the
Army, Rapid Warning Office, Fort Detrick,
Frederick, Md., developed a number of immunofluorescence systems for the specific detection of
biological agents in air samples. Certain techniques developed during that time appeared applicable to the automation of current, manually
conducted immunofluorescent diagnostic tests.
A survey of such tests was made which led to the
selection of a first candidate test for automation.
Factors considered in the survey included costs
associated with the tests, levels of skill required to
conduct the tests, total number of tests conducted,
and the need or desirability to increase the number
of tests that could be performed.
The fluorescent treponemal antibody-absorption (FTA-ABS) test (2) was selected as the first
861
candidate for automation primarily because it
appeared to us that there was a need for a syphilis
screening test more sensitive and specific than the
present, extensively used Venereal Disease Research Laboratory (VDRL) slide test. It is estimated that there were approximately 38 million
standard tests for syphilis (STS) conducted in
1968 in the United States, of which approximately
90 to 95% were the VDRL slide test. Data reported (1) indicate that correlation of test results
with diagnostic categories is 77% for the VDRL
and 95%0 for the FTA-ABS. The cost and time
required to conduct the FTA-ABS test have prevented its use for large-volume testing; however,
it appeared to us that, if automated, it might be
found suitable for such screening, thus permitting
more accurate case finding and diagnosis and subsequent improvement in control of the disease.
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
Received for publication 14 August 1969
862
BINNINGS ET AL.
APPL. MICROBIOL.
showed that the agreement between the manual
and automated FTA-ABS tests in this study was
87%. Several changes were made in the AFTA
test procedure so that it would more nearly approximate the manual procedure. Subsequent test
agreement on 201 specimens in clinically defined
categories increased to 90%, with the manual
FTA-ABS test showing slightly higher sensitivity.
Additional refinements are being made which are
expected to raise the sensitivity of the AFTA test
and to improve the agreement with the manual
procedure (G. W. Stout et al., American Public
Health Association Annual Meeting, Detroit,
Mich., 1968).
Simultaneously with this test program, 15 instruments of the production design were fabricated, and a program was formulated by the
National Communicable Disease Center for
further field evaluations at California State Department of Public Health Laboratory, Berkeley;
Venereal Disease Research Laboratory, National
Communicable Disease Center, Atlanta, Ga.;
Georgia State Department of Public Health Laboratory, Atlanta; and New Jersey State Department of Public Health Laboratory, Trenton. As
of this writing, the Berkeley and National Communicable Disease Center programs are completed and the Georgia and New Jersey programs
have been initiated. The results of these test
lished (3, 4).
Under phase II of the program, 1 December programs will be the subject of individual papers
1967 to 9 August 1968, a "breadboard" instru- to be published by each testing agency.
ment was developed and tested. A prototype
INSTRUMENT
model was established and four prototype instruThe automated instrumentation, known as the
ments were designed and fabricated. The prototype systems were tested at Aerojet-General SeroMatic System, is comprised of two major
Corp., at the Venereal Disease Research Labora- items: a processor (Fig. 1) and a microscope attory, National Communicable Disease Center, at tachment (Fig. 2).
The processor is an electro-pneumatically conSan Bernardino County Health Laboratory, San
Bernardino, Calif., and at Long Beach Memorial trolled laboratory test instrument which autoHospital, Long Beach, Calif., during the period 26 mates the performance of the FTA-ABS test on as
January 1968 to 9 August 1968. A total of 4,500 many as 200 test specimens per day (7.5 hr of
sera were tested. Results were utilized to make continuous processor operation).
functional improvement during the test program.
Specimens are processed sequentially, the first
Phase III of the program, 10 August 1968 to 21 specimen being fully processed and ready for
October 1968, comprised the application of ex- microscope examination 85 min, 45 sec after
perience from phase II into a production design processor startup. Subsequent specimens are fully
and the fabrication of one production unit. This processed at intervals of 1 min, 45 sec thereafter.
unit was then used in a study by the Venereal Figure 3 shows the process sequence, and Fig. 4
Disease Research Laboratory, National Com- and 5 are internal views of the processor.
The processor is provided with two reservoirs
municable Disease Center. In the study, the automated fluorescent treponemal antibody-absorp- for containing the wash fluids used during specition (AFTA) test was performed on 859 speci- men processing and with a set of 20 reusable 10mens from clinically defined donor groups, and cavity serum-sorbent applicator blocks.
Processor specification. The processor is 35
the results were compared with those of the
manual FTA-ABS test. The two tests were per- inches wide X 25.5 inches deep X 22 inches high
formed independently by different groups of (89 X 65 X 56 cm) and weighs approximately 120
serologists at the Venereal Disease Research lb (54.5 kg). Utility requirements are electrical
Laboratory. A comparison of the results obtained power (120 v, 60 Hz at 10 amp maximum) and air
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
Even with present patterns of use of serological
tests, an automated version of the FTA-ABS test
would provide a more ready means for detecting
patients having false-positive reactivity in cardiolipin tests for syphilis and would enhance the
detection of patients with primary, latent, or late
syphilis [clinical stages of the disease in which
present routine cardiolipin tests are not as sensitive as the FTA-ABS test (2) ].
DEVELOPMENT PROGRAM
A development program for the automation of
the FTA-ABS test was undertaken in 1967 by
Aerojet-General Corp., with technical assistance
and liaison from the Verneral Disease Research
Laboratory, National Communicable Disease
Center, Atlanta, Ga.
During the first phase of the program, 3
January 1967 to 30 November 1967, the manual
FTA-ABS test was studied to determine the effects
of variations in protocol and the quality of commercially available reagents, and to develop techniques for reagent production. From this study, a
protocol for the automated test was established, a
process for the production of high quality reagents
was proven in pilot lots, and simulated automated
tests were conducted successfully. Certain aspects
of the reagent development work have been pub-
AUTOMATED INSTRUMENT FOR FTA-ABS TEST
pressure (40 to 50 psig at 2 ft3/min). A floor sink
or 10-gal (38-liter) reservoir is required for collection of process fluid waste. Acceptable ambient
temperature and relative humidity limits are 20 to
30 C and 30 to 90%, respectively.
863
Reagents required to perform the automated
FTA-ABS test are as follows. Antigen: one REDIFIX antigen slide per specimen. (The slide is
comprised of a 25 X 75 x 1 mm thick styrene
plastic frame into which is mounted a 22 X 40 mm
FIG. 1. Processor.
FIG. 2. Microscope attachment.
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
VOL. 18, 1969
864
BINNINGIs ET AL.
APPL. MICROBIOL.
no. 1/2 cover glass. On the cover glass is a meth- specimen. PBS:approximately 20 ml per specianol-fixed 2.5-mm spot of Nichols T. pallidum men. Distilled water:approximately 160 ml per
antigen. After fixing, the antigen density is 50 to specimen.
Process sequence. The proces sequence is
100 organisms per high power field.) Sorbent:
0.1 ml per specimen. Patient or reference serum: graphically illustrated in Fig. 3 and is described
not less than 0.025 ml per specimen. Titered anti- below by following one slide through the entire
human gamma globulin conjugate labeled with sequence.
At the load station (station no. 1), a REDI-FIX
fluorescein isothiocyanate [diluted in 2% Tween
80 in phosphate-buffered saline (PBS)]:0.1 ml per antigen slide is inserted, by the loader mechanism,
PBS WASH -10 SECONDS
PBS SOAK -7 MINUTES
1ST D/W WASH 30 SECONDS
-'
1st DRY -20 SE
APPLY
%
CONJUGATE
%
'FINAL DRY -7 MINUTES
*2nd D/W WASH - 30 SECONDS
TOTAL PROCESSING TIME
-
85 MIN., 45 SEC.
FIG. 4. Internal top view.
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
FIG. 3. Processing sequence.
VOL. 18, 1969
AUTOMATED INSTRUMENT FOR FTA-ABS TEST
865
from a slide-handling magazine into one of 50
slide-holding positions which are equally spaced
around the periphery of a circular rotating table.
The rotating table is advanced (indexed) 1/50
revolution (1 station) counter-clockwise every 1
min, 45 sec. Subsequent slides are inserted into
adjacent slide-holding positions as these positions
are moved into line with station no. 1.
At 3.5 min after a slide has been inserted into a
slide holder, it becomes positioned under the
serum-sorbent applicator (station no. 3) where the
serum-sorbent mixture is forced from a cavity of a
serum-sorbent block onto the antigen spot on the
slide. This station is provided with a slide-sensing
interlock to maintain synchronization of serum
specimens with slides.
The slide is then indexed through the first incubator. The first incubator is 17 stations in
length; therefore, the slide will remain in the
incubator 29 min, 45 sec, during which time the
antigen-serum reaction takes place. Incubator
temperature is maintained at 37 C.
The first incubation is terminated by indexing
the slide out of the incubator and under the PBS
wash nozzle. This nozzle applies PBS, pH 7.2,
onto the slide for a period of 10 sec. Upon termination of the wash, a portion of the PBS wash
fluid remains on the slide and in contact with the
serum-reacted antigen spot. During the remainder
of the time which the slide remains at this station,
as well as during the time it spends at the next
three stations (approximately 7 min total), PBS
soak occurs.
When the table is next indexed (this now being
42 min from the time the slide was inserted into its
holder), the slide is brought to station no. 25
where the slide is deflected downward and rinsed
with distilled water for a period of 30 sec to remove the PBS and the particulate matter not
related to the antigen-serum reaction. Upon termination of the distilled water wash, the air dry
nozzle (also located at this station) is activated for
20 sec to remove all distilled water from the top
surface of the slide cover glass.
Next, the table is indexed to the conjugate
applicator station where a 0.1-ml drop of conjugate is applied to the antigen spot on the slide
cover glass.
The slide advances through the second incubator, remaining in the incubator for a period of 29
min, 45 sec. Incubator temperature is maintained
at 37 C.
The slide is next indexed to the second distilled
water wash station (station no. 44), where a 30sec distilled water wash is applied to remove residual conjugate.
The slide is then indexed through four stations
where large droplets of distilled water are wiped
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
FiG. 5. Bottom view.
866
BINNINGS ET AL.
APPL. MICROBIOL.
from the side and warm air from the final dry functions are interlocked to occur during a given
blower is forced across the slide to remove final timer cycle only if a slide is present at the station
traces of moisture.
where the function is to occur.
The slide is finally indexed to the unloader staFigure 6 graphically illustrates the operation of
tion (station no. 49) where the slide is ejected from all timer-controlled functions where the "home"
the rotating table onto an unload platform. The position of the cam timer is specified as reference.
unload mechanism carries the slide from this plat- Those functions shown as solid lines occur each
form into a slide unload magazine.
time cycle. Those functions shown as dashed lines
The process sequence is terminated (85 min, 45 occur only when a slide is present at the stations
sec after the slide was inserted into a slide holder), where such functions are to occur.
and the slide is ready for examination with a
The timer is of a design that permits rapid and
fluorescence microscope assembly.
simple alteration of the timer cycle program,
Processor control. All time-related operations where desired. This includes changing the point in
are controlled by a rotating cam timer which is time, during a timer cycle, that a function occurs
attached to the underside of the instrument deck and, in the case of the three washes, the duration
and accessible through a hinged panel on the front of the washes.
face of the instrument.
Microscope attachment. The microscope attachThe cam timer is driven by a 1 rev/min syn- ment, when installed onto a dark-field fluorescence
chronous motor and is appropriately geared to microscope, permits REDI-FIX antigen slides to
rotate one full revolution every 1 min, 45 sec.
be visually examined at a rate consistent with the
A total of 14 cams are attached to the timer maximal daily capacity of the processor by minishaft. Eleven of these cams are operational, the mizing the handling of slides and minimizing the
remaining three are spares. As the shaft rotates, necessary manipulation of microscope controls by
each cam either opens or closes an associated elec- the operation. Figure 7 shows a typical installatrical switch or pneumatic valve at a preset time tion of the attachment onto a microscope. [The
and for a preset duration. These switches and following combination of microscope equipment
valves, in turn, control the various mechanical has been found suitable for use in examination of
functions of the processor.
processed slides: illumination system consisting
Certain functions of the process occur during of an Osram HBO-200 ultraviolet light source and
each revolution of the timer cycle, such as table 6-v 30-w tungsten light source; excitation filtration
indexing and slide loading. Serum-sorbent appli- consisting of a 2-mm thick KG 1 (heat absorpcation, fluid washes, and conjugate application tion) filter, a 2-mm thick BG-12 (or equivalent)
L
CAM
ONE REVOLUTION OF SEQUENCE PROGRAMMER (ONE CYCLE)
0
FUNCTION
A TIMER HOME SWITCH (ELECT.)
B TABLE INDEX
C SLIDE LOAD/UNLOAD
D SLIDE EJECT
E SERUM INDEX
F SERUM DROP
G CONJUGATE DROP
H 1st AIR DRY
1st DISTILLED WATER WASH
J 2nd DISTILLED WATER WASH
K SPARE
1
L SPARE
M SPARE
N PBS WASH
L
LEGEND:
...J
OPERATES EVERY CYCLE
_ _
FIG. 6. Timing control diagram.
-L
_ OPERATES ONLY IF
INTERLOCK ENGAGED
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
TIME (SEC)
VOL. 18, 1969
AUTOMATED INSTRUMENT FOR FTA-ABS TEST
867
filter and a 2-mm thick BG-38 (or equivalent)
filter; a 2-mm thick OG-1 (or equivalent) barrier
filter, an immersion-type dark-field condenser
having a numerical aperture of 1.20; a 40 x high
dry objective with a numerical aperture of 0.65;
and a 10 x eyepiece.]
The attachment accepts a processor unloader
magazine containing the processed slides. The
slides are accurately positioned under the objective lens of the microscope by the operator. The
design of the attachment is such that the antigen
spot on the slide is in sufficient focus to require
only very slight adjustment of the microscope fine
focus control to obtain a sharp field image.
A glycerol feed assembly is provided with the
attachment (see Fig. 7). When a slide has been
positioned under the objective, the operator depresses and releases the plunger of this assembly
to glycerol-couple the slide cover glass to the darkfield condenser.
After examining a slide, the operator positions
the next slide under the objective, thereby pushing
the previous slide clear.
The unique feature of this method is that the
REDI-FIX slides are placed in the attachment in
an inverted position; that is, the reacted antigen
spot is on the lower side of the slide. By handling
slides in this manner, the need for installing a
cover glass over the specimen (as with conventional slides in fluorescence microscopy) is eliminated.
DISCUSSION
The feasibility of utilizing the automated instrumentation in the performance of routine immunofluorescence tests has been demonstrated. This
has been accomplished by successfully automating
the repetitious, time-consuming, and often tedious
steps, including those necessary during slide readout, associated with manually performed tests.
Test results to date give general indication that
the FTA-ABS test has been successfully automated. The subject instrument will permit an
increase in the total number of tests which can be
performed per unit operator time to a point where
the test may be used more extensively for screening for syphilis.
Preliminary experiments (Norins et al., to be
published) have demonstrated the feasibility of
using this instrumentation in the detection of
antibodies to malaria parasites, toxoplasma organisms, and nucleoprotein (antinuclear factor).
Experimentation continues in an effort to discover additional immunofluorescence tests which
can be similarly automated.
It is felt that the inherent functional flexibility of
this instrument may be a further measure of its
value, providing investigators with a fertile tool in
the field of developmental research.
ACKNOWLEDGMENTS
The instrument development program was inspired by Leslie C.
Norins of the Venereal Disease Research Laboratory, National
Communicable Disease Center, Atlanta, Ga. In addition, G. W.
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
FIG. 7. Microscope with attachment and glycerol feed assembly.
BINNINGS ET AL.
Stout, V. Falcone, A. Wallace, W. Duncan, and J. Lewis of the
Venereal Disease Research Laboratory provided valuable advice
and assistance in the conduct of evaluation tests and furnished information helpful to the refinement of the design. Others who
rendered valuable assistance in conducting evaluation test programs and making suggestions which were helpful for the improvement of the design were R. Wood, E. Coffey, and R. Jue, California
State Department of Public Health Laboratory, Berkeley; S.
Brandon, C. Gates, and J. Dalton, Georgia State Department of
Public Health Laboratory, Atlanta; E. Thomas and C. Levy, New
Jersey State Department of Public Health Laboratory, Trenton;
F. Gettman and H. Davis, San Bernardino County Health Department Laboratory, San Bernardino, Calif.; and M. Mercer,
Long Beach Memorial Hospital, Long Beach, Calif. We thank E.
Mishuck of Aerojet-General Corp. for encouragement and counsel
throughout the course of these studies.
APPL. MICROBIOL.
LITERATURE CITED
1. Deacon, W. E., J. B. Lucas, and E. V. Price. 1966. Fluorescent
treponemal antibody-absorption (FTA-ABS) test for syphilis. J. Amer. Med. Assoc. 198:624-628.
2. Hunter, E. F., L. C. Norins, V. Falcone, and G. W. Stout. 1968.
The fluorescent treponemal antibody-absorption (FTAABS) test, development, use and present status. Bull. World
Health Organ. 39:873-881.
3. Roberts, M. E., J. N. Miller, and G. F. Binnings. 1968. Reduction of nonspecific background staining in the fluorescent
treponemal antibody-absorption test. J. Bacteriol. 96:15001506.
4. Roberts, M. E., J. N. Miller, T. C. Pringle, and G. F. Binnings.
1968. Shelf life of fluorescent treponemal antibody-absorption test reagents. J. Bacteriol. 96:1507-1511.
Downloaded from https://journals.asm.org/journal/am on 11 August 2021 by 3.238.203.22.
868