zy
Lasers in Surgery and Medicine 10510-523 (1990)
Limitations of a Thermal Camera in
Measuring Surface Temperature of
Laser4rradiated Tissues
zyxwvu
Jorge H. Torres, MD, Thomas A. Springer, MS, Ashley J. Welch,
John A. Pearce, PhD
PhD,
and
zyxw
Biomedical Engineering Program, University of Texas at Austin, Austin 78712
Thermal cameras are used in research laboratories to measure
tissue temperature during laser irradiation. This study was an
evaluation of the accuracy of a 3-5 pm thermal camera and two
S 1 2 pm cameras in detecting the maximum temperatures of
small targets. The size of the targets was within the range of laser
spot diameters which are used for vessel welding, angioplasty,
and dermatology. The response to a sharp thermal edge was measured and analyzed for the three cameras, which had a scanning
rate of 30 frames per second. The response of the 3-5 pm camera
to reference black body targets of different sizes was also studied.
It was found that the detector system required an average of
2.44 p s to reach 90% of maximum step response for the S 1 2 pm
system and 5.85 ps for the 3-5 pm system. With a 3 x telescope
and a 9.5 inch focal distance close-up lens, the 3-5 pm camera
underestimated the temperature of targets smaller than 2.0 mm
because of its slow detector response. Although the 8-12 pm camera provides more accurate measurements due to its faster detector response, it still underestimates the temperature of targets
smaller than 900 pm, when similar magnification and focal distance are used.
Methods to compensate for the inaccuracies are discussed, including empirical correction factors and the inverse filtering
technique.
zyxwvut
zyxwvu
Key words: infrared camera, tissue temperature, measurement error
INTRODUCTION AND BACKGROUND
The cameras measure temperature by detecting the infrared radiation emitted by the obThermal cameras have been used in industry jects in their field of view. Objects radiate energy
to measure relative temperatures of specific ma- according to their absolute temperature and in
terials, and by the military to identify targets. proportion to their surface emissivity. An ideal
They have also been used clinically to help in the
diagnosis of breast cancer or t o study circulatory
problems. These cameras are now becoming more Accepted for publication August 1, 1990.
common in laboratories established for biomedi- Address reprint requests to Dr. Jorge H. Torres, Biomedical
cal research. In the particular area of laser appli- Engineering Program, Univ. of Texas at Austin, ENS 610,
Austin, TX 78712.
cations t o medicine, thermal cameras are used to
measure tissue surface temperature during laser Dr. A.J. Welch is the Marion E. Forsman Centennial Professor of Electrical and Computer Engineering and Biomedical
irradiation. The information obtained is corre- Engineering.
lated with the extent of damage observed histowork was supported in part by the Free Electron Laser
logically, and provides insight into the kinetics This
BiomedicaUMaterials Science Program-ONR contract numand thermodynamics of laser interaction with tis- ber N00014-86-K-0875-and in part by the Albert and Clemsue.
mie Caster Foundation.
0 1990 Wiley-Liss, Inc.
z
zyxwvu
Limitations of a Thermal Camera
51 1
black body, a body which absorbs all radiation
incident upon it and reflects or transmits none,
Detector
has an emissivity of 1.0. It also emits the maxi0
mum possible amount of energy at any specific
I
temperature and wavelength. Most bodies, however, may be described as “gray bodies” with
emissivity less than one and approximately indeCamera
pendent of wavelength. For example, some polished metallic surfaces have an emissivity as low
Target
as 0.05, whereas vascular tissues have emissivities from 0.93 to 0.98 [ll.
For a “gray body,” the total emissive power
(over all wavelengths) is [2]:
Fig. 1. a: Detector instantaneous field of view (IFOV = p).
I
m
zyxw
zyxwvut
zyxwvu
zyxwvutsrqponm
E(T)
=
eaT4
(1)
where (T is the Stefan-Boltzmann constant,
5.67 X lo-’ (W/m2.K4),and E is the emissivity (0 5
E I11,which is defined as the ratio of the radiant
energy emitted by an object at a specific temperature to the radiant energy emitted by a black
body at the same temperature.
Two infrared spectral bands exhibiting minimum absorption by the atmosphere are used for
thermographic imaging: 3-5 pm and 8-12 pm.
The detector for the 3-5 pm band is normally a
crystal of mercury-cadmium-telluride (HgCdTe)
or indium antimonide (InSb). Only a crystal of
HgCdTe may be used as a detector for the 8-12
km band owing t o its narrow energy band gap.
Due t o the low photon energy of the infrared radiation t o be measured, it is necessary to cool the
detector with liquid nitrogen t o -196°C (77 K) in
order to minimize thermal noise. In a typical instrument, the electrical conductivity of the detector changes proportionally with the total power of
the radiant energy received. A voltage is obtained
from a resistance bridge configuration and translated into gray level intensity or color band displayed on a monitor.
At each instant of time, the detector can
“see” only a certain region which falls within its
“instantaneous field of view” (IFOV) or acceptance angle. The IFOV and the distance between
the detector and the target determine the spatial
resolution of the camera. The smaller the IFOV
the better the detector can discriminate between
two points, and therefore the better the resolution. To form a complete thermal image, the detector must scan an entire scene. This is accomplished by mechanically scanning the detector
either by rotating prisms or oscillating mirrors,
resulting in a “flying spot scanner.” The IFOV
determines the size of the individually resolvable
The diameter of the viewed area w is determined by the IFOV
and the distance to the detector D. b Optical path followed by
the emitted radiation to reach the detector (Det) inside the
camera.
samples or independent picture elements. The
definition of IFOV is presented graphically in
Figure la, and is expressed in the equation:
IFOV
=
p
=
2tanp1[w/2D1
(2)
where w is the width of the region to which the
detector responds at any given instant of time,
and D is the distance from the target t o the detector. The distance D includes the optical path
internal to the camera as shown in Figure lb. Due
to focusing lenses and other optics, an effective
distance D’ must be determined for each particular scanner.
When the thermal camera is focused on an
opaque surface (with zero transmittance), the radiation (radiosity) received by the detector is:
where J,(Ts,Te,E) is the measured radiosity,
Eb(Ts) is the black body emissive power at the
temperature of the surface Ts, and Eb(Te) is the
black body emissive power at the temperature of
the environment Te, all given in Watts/m2. The
surface reflectance p is related t o the emissivity
by p = 1 - E. Knowing the radiosity measured by
the camera, the temperature of the environment,
and the emissivity of the surface, Eb(Ts) can be
obtained from equation (3). Once Eb(Ts) is found,
the surface temperature Ts can be estimated from
calibration tables which use Planck’s law of spectral distribution of energy as a function of temperature [21.
Since the relationship between emissive
512
zyxwvutsrqp
zyxwv
zyxwvutsrqponmlkjih
Torres et al.
1500 1
0
50
100
a color monitor. Initial experiments performed
with this camera provided tissue temperatures
lower than expected during irradiation with
small spot size laser beams. In vessel welding experiments using a C02 laser and a spot size of 500
pm, the camera indicated temperatures considered too low for the macroscopic and histologic
damage observed. Experiments with argon laser
irradiation of human aorta using spot sizes between 700 pm and 1 mm led to a similar observation. Also, there have been some reports on COz
laser irradiation of dog aorta and rat femoral and
carotid arteries at spot sizes of 200 t o 350 pm
[5-71 in which the authors, although using different thermographic imaging systems, have measured temperatures that, in our judgment, may be
lower than the actual values. Therefore, we decided to carry out the present study to evaluate
the accuracy of thermal cameras in detecting the
temperature of targets smaller than 2 mm. In this
paper an experimental evaluation of the system
response to small targets is presented, and methods to correct for errors are discussed.
zyxwvutsrqpo
zyxwvutsrqpon
zyxwvutsrqp
150
200
250
300
Blackbody Temperature ("C)
Fig. 2. Non-linear relationship between band-limited emissive power and black body temperature. (Data were obtained
from Ryu L121.1
power and temperature is not linear (31, and since
factors like atmospheric absorption and attenuation by filters and other optics in the camera affect the measured radiosity, calibration must be
performed for each particular camera in order t o
obtain the correct target temperature. The nonlinear relationship between band-limited emissive power and black body temperature for both
the 3-5 and the 8-12 pm bands is presented in
Figure 2. A table with values corresponding to the
curves in Figure 2 can be constructed. For speed
in computation, the curves can be approximated
by polynomial functions [41. In general, the Calibration method uses two or more reference black
bodies at known temperatures whose emissive
powers are measured by the camera. Calibration
tables are then utilized to estimate the target
temperature.
In the past few years, in our laboratory, we
have used an 8-12 pm camera to measure tissue
surface temperature during laser irradiation,
with spot sizes usually varying from 700 pm to
2 mm. We calibrated the thermographic images
by placing two reference black bodies at known
temperatures in the scene close t o the target. The
gray scale thermal images were recorded on videotape and subsequently digitized and analyzed
on a computer. Using the gray level and temperature values of the black body references as well
as the calibration tables [4], the temperature of
the target was calculated from its measured gray
level (linearly related to its emissive power).
More recently however, we acquired a new
3-5 pm thermal camera which uses internal calibration t o calculate and display temperatures on
MATERIALS AND METHODS
Thermographic Imaging System
A 3-5 pm thermal camera model 600L from
Inframetrics, Inc., was primarily used for this
study. Two 8-12 pm cameras, models 525 and
600L from Inframetrics, Inc., were also tested in
order t o compare detector responses. All three
cameras use oscillating mirrors to scan the scene
horizontally and vertically and produce images at
30 frames per second.
The thermal cameras tested in this study use
two mirrors oscillating at 3.933 kHz and 60 Hz to
scan the scene horizontally and vertically, respectively. Like standard video, the vertical scanning
follows a sawtooth waveform and images are produced at 30 frames (or 60 fields) per second. However, the horizontal scanning has a uniform oscillatory pattern instead of the sawtooth form of
standard video. Due to the lower frequency and
the oscillatory type of the horizontal scanning,
each line is scanned in 125 p s (instead of 53 p s of
standard video) and two consecutive lines in a
field are scanned in opposite directions. The result is that only 7,866 horizontal lines are
scanned every second (as opposed to the 15,750
lines of standard video). There are 100 horizontal
lines of actual infrared information per field (or
200 per frame), and each line contains 256 digital
samples [8,91. The differences between standard
zyxw
z
zyxwvutsrqponml
zyxwvuts
Limitations of a Thermal Camera
4
53 p e c
>
<
125 p e c
513
>
Black Body
Thermocouple
Thermal Camera
Standard Video
Aluminum
Shield
Thermal Camera
(with oscillating mirrors)
Fig. 3. Comparison between standard video raster and thermal camera raster. For the thermal camera, horizontal scanning has a uniform oscillatory pattern instead of a sawtooth
form typical of standard video.
video and thermal camera scannings are shown in
Figure 3.
A computer digitizes the analog signal from
the camera and then displays it on a TV monitor.
In order to match video standards, the digital information is loaded into a FIFO (First In First
Out) memory for the right-going scan, and into a
LIFO (Last In First Out) memory for the leftgoing scan. While the camera scans one line, the
information stored from the previous line is read
twice out of memory and displayed twice on the
TV monitor in order to generate the 15,750 lines
per second of standard video. The final video image contains 400 lines of infrared information
with 256 pixels per line. The rest of the 525 TV
lines are used for text and gray scale display.
Since 256 digital samples are taken during
the scanning of one horizontal line and 7,866 lines
are scanned every second, the camera bandwidth
is:
Bandwidth = 7,866 linedsec X 256
sampledine = 2.01 MHz.
The cameras have a total horizontal field of view
of 18"t o 20" (314 to 350 milliradians). The instantaneous field of view (IFOV) is 2mRad for the
8-12 pm cameras and 3.5 mRad for the 3-5 pm
camera [8,9]. Therefore, although sampling is
performed 256 times during a horizontal scan,
there are only 175 (350 mW2 mR) independent
samples in a line for the 8-12 pm system, and 100
(350 mW3.5 mR) for the 3-5 pm system.
The manufacturer specifies a minimum detectable temperature difference of 0.1"C for the
8-12 pm cameras and 0.2"C for the 3-5 pm camera [S]. Each of the three cameras uses a mercurycadmium-telluride detector which has a specified
time constant of 0.5 t o 1 ps [lo].
zyxwv
Black Body
Aluminum Shield
Fig. 4. Experimental setup to measure the response to a thermal step.
Experimental Evaluation
A) System response to a thermal step.
For small targets, the accuracy of the temperature measurements depends on the detector IFOV
and slew rate. These parameters were assessed by
measuring the detector response t o a sharp
change in temperature or step during horizontal
scanning. A temperature step was created by taking a 2 inch circular black body and covering half
of its surface with a shield of aluminum foil as
shown in Figure 4. In this way, during horizontal
scanning, the detector initially saw "room" temperature reflected by the aluminum foil, and at
the edge of the foil it saw a sudden, sharp change
to the temperature of the black body which was
electrically heated to a known temperature. The
edge of the aluminum foil was made as sharp and
straight as possible t o avoid blurring effects. Each
camera was placed at 9.5 inches from the black
body and focused t o obtain the sharpest image of
the edge. The images were recorded on videotape
in gray scale for further analysis. No telescope or
external optics were used with the cameras. One
hundred percent field of view was used (no zoom),
providing a total horizontal view of approximately 11 cm. The response was evaluated for
three temperature steps: 25-36.2"C, 25-52.4"C,
and 35-105°C. Images were digitized on an Intel
Computer iSBC 86/30 and values of gray level
were obtained for each pixel in a selected subframe centered at the edge. Minimum and maximum gray levels were determined by averaging
the value of 10 pixels on a particular line in the
cold area before the step and in the hot area after
zyxwvutsrqp
zyxwv
Torres et al.
the step respectively. Several lines were analyzed Determination of the Camera Point
in order to find a typical response curve for both Spread Function
514
right-going (cold-to-hot) and left-going (hotto-cold) scans. During the experiments, an oscilloscope was used to corroborate detector rise time
and curve shape.
In order to evaluate the effects of using a
telescope and a close-up lens on the detector response to the sharp thermal step, another set of
experiments was performed with the 3-5 pm
camera. The camera was used with a 3~ telescope and a 9 inch close-up lens and was focused
at 9.5 inches (24.13 cm) from the sharp edge. The
25-52.4”C thermal step was used in this case.
Computer analysis similar to that described
above followed the experiments.
B) Camera temperature readings for
small targets. The accuracy of the 3-5 pm camera in measuring the maximum temperature of
small black body targets was evaluated by placing a variable slit in front of a black body with a
1.6 mm circular aperture. This camera had internal calibration and provided a temperature reading for the target. The black body was electrically
heated and a thermocouple measured the temperature of its cavity. The width of the slit could be
varied from 0.1 to 1.6 mm. A 3 x telescope and a
9 inch close-up lens were used. The camera was
placed at 9.5 inches from the slit so that the best
focus was obtained. The black body was maintained at 100°C for easy reference. For each width
of the slit, the maximum temperature read by the
camera was recorded. To compare with these experiments, two black bodies with sizes of 0.5 and
1.0 mm respectively were made by drilling two
holes of corresponding diameters in an aluminum
block. The holes were deep enough (16mm) to
make the cavities approximate ideal radiators.
The surface of the aluminum block was highly
polished. The block was heated slowly by a hot
plate, while thermocouples placed at the ends of
the cavities measured the corresponding temperatures. The temperatures read by the 3-5pm
camera were compared with those measured by
the thermocouples. In addition, temperature
readings for a separate black body with a 2 mm
diameter were also tested.
To evaluate the system response as a function of target temperature, the black body size
was kept constant at 0.5, 1.0, and 1.5 mm (fixed
slit width) while varying the temperature from
30” to 150°C. Target temperatures and the corresponding temperature values measured by the 3 5 pm camera were recorded and compared.
The image degradation effects caused by a
particular imaging system (in this case a thermal
camera) can be assessed by measuring the point
spread function (PSF)of the imaging system. The
PSF is the response of the system to a point source
input. Considering that the image is a two-dimensional distribution of point sources, knowledge of
the PSF can be used to determine the action of the
system on the image [lll. For a thermal camera,
the detector response as well as the scanningmechanism and the optics involved produce a degradation action or blurring operation on the input
picture. Calling this blurring operation (i.e., point
spread function) h(x,y), the input image f(x,y),
and the output (degraded) image g(x,y), then:
where * indicates 2-D convolution. If the input
picture is only a hot point source, the output will
be the PSF of the camera.
In practice, for a thermographic imaging system with limited spatial resolution, the dimensions of a hot point source which could be used to
determine the PSF are difficult to define. However, the following method can be used to find the
point spread function h(x,y) of the camera [4,111:
The system response to an edge or thermal step
can be measured in the way described in the previous section (see Fig. 4). The system response t o
a sharp edge oriented in a particular direction is
defined as the “edge spread function” (ESF) for
that particular direction or h,(x,y). The derivative
of the ESF in any direction is the “line spread
function” (LSF) in that direction or hL(x,y). In
other words, the LSF is the system response to a
line oriented in the direction of the edge described
above. The projection of the LSF is basically a
point. Experimentally, the thermal edge can be
rotated 360” in small steps (which depend on the
precision wanted) and the ESF can be measured
for different orientations. The corresponding
LSFs can be then calculated. The projections of
the LSFs taken at different orientations (defined
by an angle 8 + 90”) allow the two-dimensional
reconstruction of the point spread function. See
Figure 5 for a graphical summary of this process.
The reconstruction of the PSF from the projections of the LSFs is based on the “Fourier Slice
Theorem” ill3 which states that the one-dimensional Fourier transform of the projection of the
zyxwv
zyxw
-
zy
zyxw
metry, this final LSF was rotated 360" to obtain
the camera PSF.
+
zyxwvutsrqp
zyxwvutsr
zyxwvuts
Gaussian Vs. Uniform Targets
A
Edge Spread
Function (ESF)
Line Spread
Function (LSF)
Point Spread
Function (PSF)
Fig. 5. Method to obtain the point spread function from the
projections of the line spread functions a t different orientations.
image function (in this case the PSF) along a line
s oriented at 0 + 90" is equal to the two-dimensional Fourier transform of the function along a
line t oriented at an angle 0 from the horizontal
axis x (see Fig. 5). A reconstruction algorithm
like the filtered-backprojection method [ 111 can
be used. In this study however, a more simplistic
approach was taken to get the PSF for the 3-5 pm
thermal camera. First of all, due t o the way the
scene is scanned and the way the data are stored
and displayed, the PSF is asymmetric. Despite
that fact, since the detector response during the
horizontal scanning is the most limiting factor for
the peak temperature measurements, only the
ESF and the LSF for 0 = 0 (vertical edge) were
measured in this study. Then, the PSF was constructed assuming circular symmetry by rotating
360" the projection of the LSF for 0 = 0 (see also
Springer et al. [121). The PSF indicates the blurring effect of the camera on a thermal image.
Once the PSF is known, restoration (to a certain
degree) of an original thermal image is possible
from the camera output. In this investigation, the
PSF of the 3-5 pm camera model 600L was measured. Ryu [41 had measured the PSF for the 8-12
pm camera model 525, also tested in this study.
Since the response of the 3-5 pm camera to
a cold-to-hot step (right-going scan) was different
from its response to a hot-to-cold step (left-going
scan), as will be shown later, the following procedure was used to establish a PSF for this camera:
Ten lines were selected for each scan direction.
For each line, median filtering was applied to remove the noise, and then the edge spread function
(ESF) was differentiated to obtain the corresponding line spread function (LSF). The ten LSFs were
averaged together t o produce one LSF per scan
direction. The two LSFs, one for the right-going
scan and one for the left-going scan, were again
uniform (pulse type) temperature distribution
across the cavity. However, before heat conduction becomes significant, most temperature rises
in tissue during laser irradiation have a gaussian
distribution (which has smoother edges and is not
uniform across the spot). In order to compare the
system response to gaussian targets with the response to uniform targets of corresponding diameters, the following procedure was used:
Targets with uniform and gaussian temperature distributions were simulated on a computer, and the PSF previously obtained for the
3-5 pm camera was used to calculate the camera
response to the peak temperature of the targets.
For each target type, the size (diameter) in pixels
was varied. The width of the uniform rectangular
targets was taken as the diameter. For the gaussian targets, the diameter was the distance between the points at which the temperature
dropped to l/e2of the central value. Once the onedimensional temperature distribution was simulated, a fifth-degree polynomial approximation to
the 3-5 pm band-limited power vs. black body
temperature curve was used to obtain the emissive power distribution which was subsequently
normalized to values from 0 to 100%.The normalized emissive power distribution (EJ was then
convolved with the camera one-dimensional PSF
to obtain the system response (E,) in percentage,
or: Ei * PSF = E,. For the uniform targets, system
response was assumed to be independent of the
magnitude of the thermal step, and a thermal
step from 0 to 100°Cwas arbitrarily chosen. Since
the shape of the gaussian distribution varies with
the peak temperature, three peak temperatures
were selected for comparison: 50, 100, and 150°C.
RESULTS
System Response to a Thermal Step
The detector response t o a thermal step from
25 to 36.2"C is shown in Figure 6 for the three
cameras. The graph was normalized from the average background gray level t o the maximum
gray level achieved by each camera. It represents
the percentage of the system maximum response
as a function of scanned pixels in a displayed TV
Torres et al.
516
zyxw
Large hot
object
,
,
1
2
0
3-5 pm Camera Model 600L
8-12pm Camera Model 600L
8-12 Vrn Camera Model 525
,
,
3
4
,
l
l
l
l
l
l
5
6 7 8 9 1 0 1 1
Time (psec)
Maximum.
0
objec!
zyxwvuts
zyxwvutsrq
zyxwvutsrqpon
zyxwvutsrqponmlk
l
0
A
Small hot
,
Fig. 6. Detector response to a 25-36.2”C thermal step for
three thermal cameras. Note the slower response of the 3-5
km detector.
line consisting of 256 pixels (upper scale), and as
a function of time (lower scale). In this way, the
relationship between time and pixels scanned on
a horizontal line can be observed as the system
responds t o the cold-to-hotstep. Since a line of 256
pixels is scanned in 125 ps, each pixel is scanned
in approximately 0.49 ps. As seen in Figure 6, all
three systems take some time and several pixels
in the scanning line to reach maximum response.
The 8-12 pm detectors require 2.44 p s to reach
90% of the maximum response, which corresponds
t o 5 pixels. The 3-5 pm detector system is slower
than the 8-12 pm system, requiring 5.85 p s or 12
pixels, to reach 90% of the maximum response.
In the case of the 3-5 pm thermal camera,
when the 3~ telescope and a close-up lens with
24.13 cm (9.5 inch) focal distance are inserted, the
measured horizontal viewing distance is 3.0 cm
and each pixel corresponds t o 117 pm (3.0 c m /
256). This means that the detector needs t o scan
1.4mm in the scene to reach 90% response and
2.0 mm to reach 100%response. Therefore, it will
scan through any hot object smaller than 2 mm
without reaching maximum response. The system
will then indicate a temperature value lower than
the actual temperature of the object. A graphical
representation of the problem is shown in Figure
7.
In the case of the 8-12 pm system, response
time is still a problem but it is less pronounced.
Based on Figure 6, if the 8-12 pm camera with
the same telescope and close-up lens is also focused at 9.5 inches from the target, the system
will not indicate the correct temperature for ob-
..
..
Time
Cletector
Response
(Apparent
Intensity)
Fig. 7. Sketch to show that the detector does not reach maximum response while scanning small objects.
jects smaller than 900 pm (about 8 pixels), although it will reach the 90% response rate for a
550 pm object (about 5 pixels). Numbers obtained
for both 8-12 and 3-5 pm detector responses are
presented in Table 1.
The detector response to the hot-to-cold step
or sudden temperature drop was measured by following the left-going scanning lines. Gray level
values fell from 90% t o 10% of the maximum
value in approximately 1.5 p s for the 8-12 pm
system and 2.5 p s for the 3-5 pm system. Therefore, for the 3-5 pm camera, the response to a
sudden temperature drop was more than two
times faster than the response to a sharp temperature increase: 2.5 p s (for 90-10%) vs. 5.15 p s (for
10-90%). For the 8-12 pm camera, the difference
between these two responses was much less pronounced: 1.5 p s (for 90-10%) vs. 2.0 p s (for 1090%).
During scanning, the detector response t o
rapid temperature changes in the scene depends
on two factors: the detector slew rate and its instantaneous field of view (IFOV). Looking at Figure 6, one can see that the detector response
curves have a lower slope for the first 1or 2 pixels.
The detector appears very slow to respond when it
first “sees” the thermal edge. This is actually a
blurring effect caused by the detector IFOV and is
more pronounced for the 3-5 pm camera. The effect of the IFOV is described in Figure 8. When
the detector first encounters the edge, it receives
zyxwvut
zyxwvu
Limitations of a Thermal Camera
TABLE 1. Detector Responses
Detector parameter
Time constant (0-63.2%)
Rise time (10-90%)
Time for 0-90% response
8-12 pm
system
1.6
2.0
2.44
517
3-5 pm
system
2.2
5.15
5.85
0
2
4
zyxwvutsrq
zyxw
6
8
10
12
14
16
18
20
22
24
Pixels
Fig. 9. Comparison between detector response when telescope was used and response when only standard optics were
utilized.
clF[m
__.. __.. --..-
=
-
F
zyxwvut
recall that the IFOV determines the spatial resolution of the system. When a 3 x telescope and a
close-up lens with a 9.5 inch focal distance are
used, an IFOV of 2 mRad corresponds to a distance of 200 pm and an IFOV of 3.5 mRad corresponds t o 300 pm, while a pixel displayed on the
monitor corresponds to 117 pm. In this case, the
response to a target 5250 pm is predominantly
affected by the IFOV. For targets larger than 250
pm, the detector slew rate is the most limiting
factor in the response. Mathematical expressions
for the response to an edge, based on the IFOV
alone, can be obtained for a circular window and
other geometries.
The magnitude of the thermal step did not
seem t o affect the detector response. The responses of the 3-5 pm detector to three different
thermal steps were almost identical to the response shown in Figure 6. There was basically no
difference between the curves for the 25-36.2"C
step, the 25-52.4"C step, and the 35-105°C step.
The effect of the telescope and close-up lens
on the response t o the edge was also tested. In
Figure 9, the response t o the 25-52.4"C thermal
step when the 3 x telescope and the 9 inch closeup lens were used is compared with the response
t o the same step when only standard optics were
utilized. The response was slightly slower when
the telescope and close-up lens were used, perhaps
due to some blurring of the edge by optical magnification. With the camera focused at 9.5 inches,
the entire image was magnified 3.67 times (from
a total horizontal view of 11 cm down t o 3 cm).
For small objects, of course, a telescope and a
close-up lens are required. That arrangement
zyxwvutsrq
Displayed
Pixels
Fig. 8. Description of the effect of the instantaneous field of
view (IFOV) on the response to an edge. Note the overlapping
of sampled IFOVs when they are larger than the distance
between consecutive samples.
energy from an area covering both sides of the
edge. The dimensions of this area depend on the
IFOV. The larger the IFOV the longer it takes for
the detector to "move" completely from the cold
zone to the hot one. In addition, since the total
horizontal view is 350 mRad, and since 256 digital samples are taken during the scanning of one
line, sampling is performed every 1.37 mRad.
This means that if the IFOV is larger than 1.37
mRad, overlapping of consecutive sampled IFOVs
occurs and the corresponding displayed pixels are
not independent. The result is blurring and reduced image resolution, The larger the IFOV the
worse the blurring effect. This effect is present in
the response of the cameras tested here because
the IFOV is 2 mRad for the 8-12 pm detector and
3.5 mRad for the 3-5 pm detector [91, while the
pixels displayed on the monitor correspond t o the
digital samples taken every 1.37 mRad. Therefore, for the 8-12 pm camera, totally independent
samples are displayed every 2 pixels; and for the
3-5 pm camera, completely independent samples
are displayed every 3 pixels. It is important to
zyxw
zyxwvutsrqponml
zyxwvu
Torres et al.
518
140
-0-
100
4
...................................,...........................
NoTelescope
3X Telescope +
9close-up lens
100
2ofi-,
'
zyxwvu
"
1
zyxwvutsrqpon
,
0
0
1
2
3
4
5
6
7
8
Pixels
9
,
,
1011 1 2 1 3 1 4 1 5
Fig. 10. Effect of magnifying a 0.5 mm black body with a 3 x
telescope and a 9 inch close-up lens on the intensity measured
by the 3-5 km system. The arrows indicate the slow response
at the edges due to the effect of the IFOV.
magnifies the image, improving the resolution
and allowing the detector to respond better t o
small targets, since they now include more pixels.
The improvement in the response of the 3-5 pm
camera t o a 0.5 mm black body target at 137°C
when the telescope and close-up lens are used can
be observed in Figure 10. The arrows indicate the
slower response at the edges of the black body due
to the effect of the IFOV.
Camera Temperature Readings for Small Targets
The apparent temperature measured by the
3-5 pm thermal camera focused on a black body
at 100°C as a function of black body size (slit size)
is presented in Figure 11. The camera with telescope (3x ) and close-up lens was focused at 9.5
inches from the target. It is easy to see that the
camera underestimated the temperature of the
target for all sizes up to 1.6 mm. The smaller the
target size the worse the temperature underestimation. These data confirm that the system is unable to reach its maximum response while scanning objects smaller than 2 mm. In Figure 11,
note that for a target size of 500 pm the camera
read 65°C instead of lOO"C, and for a 1.0 mm target the camera read 85°C. When the camera was
focused on the black bodies made in the aluminum block, it read 62°C and 86°C for the 0.5 and
1mm black body respectively while the cavity
thermocouples read 100°C, clearly in agreement
with the values plotted in Figure 11.On the other
hand, the camera provided the correct temperature for the 2 mm black body.
The system response as a function of target
00.0
0.2
0.6
0.4
0.8
1.0
1.2
1.4
1.6
Target Size (mm)
Fig. 11. Apparent temperature measured by the 3-5 pm
camera as a function of black body size. The black body was
maintained at 100°C.
temperature was also evaluated by varying the
temperature of a fixed size black body. The 3-5
pm thermal camera was again used to measure
the temperature of the target. A ratio of the measured temperature rise ("(3) t o the actual temperature rise ("C) as a function of the actual target
temperature above that of the environment ("C)is
presented in Figure 12a for three different target
sizes: 0.5, 1.0, and 1.5 mm. As can be seen in the
figure, this ratio increases as the target temperature is increased. The non-linear relationship
between band-limited emissive power and black
body temperature explains this behavior (see
again Fig. 2): as temperature increases, a particular emissive power difference corresponds to a
smaller temperature difference. Note that, for the
range of temperatures investigated, the ratio increases with temperature in a linear fashion. After curve fitting, linear functions with approximately the same slope were obtained for the three
cases (see the mathematical expressions in Fig.
12a). After averaging the slopes, the following relation was obtained:
z
=
b
+ 0.25 AT
(5)
where z is equal to 100 times the fraction AT measured ("C)/AT actual ("C), b is a constant which
depends on the target size, and AT is the actual
temperature rise ("C) with respect to the ambient
temperature. If, due t o some calibration error, the
camera provides a different value for the ambient
temperature when changing temperature scales,
zyxwvutsr
R”2 = 0.98
519
The value of b for the 2 mm black body (also used
in the test) was included in Figure 12b, consider100
ing that the camera provided the correct temper90
56.746 + 0.25319 AT ature at AT of 40°C.
I1
80
Therefore, for temperature rises up t o 120”C,
5x 70
34,890 o,25240 AT an empirical correction factor z has been derived
2 6 0
for the 3-5 pm camera used in this study, and
50
equations (5) and (6) can be used to estimate the
40
1.5 mm
actual temperature of targets smaller than 2 mm.
Higher target temperatures were not tested due
I.Om
to limitations in the black bodies used in our
0 0.5 mm
study. These black bodies could not be heated
10
more than 150°C. Computer simulations using
a 0 0 20 40 60 80 100 120 140
the relationship between band-limited emissive
AT actual (“C)
power and temperature (shown in Fig. 2) indicate
that the linearity presented in Figure 12a does
b = - 37.565 + 240 19s - 261 4 7 s V + 159 86sA3 - 4 8 . 1 4 7 ~ ~+45 ,767553
R”2 = 0.99
not hold at higher temperatures, and that a more
loo complicated expression for the correction factor is
90 required. However, for a limited range of temper80 atures, it is possible t o obtain a simple expression
70 for a correction factor which allows an approxiD
60mation to the true target temperature, as it was
c
done in this study for temperatures from 25°C to
50 4> 40 150°C.
The manufacturer provides a graph for a
30 function called “slit response function” (SRF),
20 which is basically the fraction AT measured/AT
10 actual for a vertical slit target of various widths
as measured experimentally [81. However, as
b 0.0 0 . 2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2
shown
above, this fraction changes with temperTarget size (mm) = s
ature and does not have a constant value. We
Fig. 12. a: Ratio of measured temperature rise to actual tem- have found that the values of the SRF indicated in
perature rise as a function of target temperature above am- the manufacturer’s manuals hold only for target
bient temperature for three different black body sizes. In all temperatures around 100°C (AT around 75°C). It
three cases, note the linear increase in this ratio with increasing temperature. b: Value of the constant b as a function of will be more accurate t o follow the procedure
target size. This value will be used in the equation z = b + shown above to derive a correction factor for a
0.25 AT, where z will be the correction factor (%).
particular thermal camera.
z = 71.041+ 0 . 2 4 2 7 0 ~ ~
N
+
2
5
:
z
zyxw
zyxwvutsr
zyxwvutsrqponmlkjih
zyxwvutsrqpon
zyxwvutsrqp
zyxwvuts
zyxwvutsrqpon
Limitations of a Thermal Camera
s.
’
0
Camera Point Spread Function
the offset must be taken into account. Based on
the curve of apparent temperature vs. target size
for a black body at 100°C (Fig. ll),the values of z
were obtained for target sizes from 0.2 t o 1.6 mm
at a AT of 75°C. Using equation (5), the corresponding values of b were obtained. These values
are plotted in Figure 12b. A fifth-degree polynomial function fitted those points extremely well,
allowing the use of the following expression t o
estimate b for a specific target size:
b
=
-37.56
zyxwvu
+ 240.19 s - 261.47 S’ + 159.86 s3
48.15 s4 + 5.77 s5
(6)
-
The PSF obtained for the 3-5 pm camera is
shown in Figure 13. In this three-dimensional figure, the x-y plane represents space divided in pixels, and the height (z direction) corresponds to the
normalized response. As indicated in the Methods
section, the line spread functions for the rightgoing scan and for the left-going scan were averaged together and the result rotated 360”.
where s is the size of the target in millimeters.
Gaussian Vs. Uniform Targets
The PSF presented in Figure 13 was used to
predict the response of the 3-5 pm system t o simulated targets with either gaussian or uniform
temperature distribution. The predicted responses
to both target types were then compared. Only
520
zyxwv
zyxwv
zyxwvutsrq
--
Torres et al.
Rectangular
Gaussian
3
0
zyxwvuts
zyxwvut
zyxwvutsrq
5
7
10
15
20
25
30
35
Diameter (pixels)
zyxwvutsr
r
n
0
.35 S 8 . 8 2 1.17
x
s
1
I
1.76
2.34
2.93
3.51
4.10
Diameter (m)
Fig. 13. Point spread function (PSF)obtained for the 3-5 pm
camera. The x-y spatial plane is divided in pixels. The height
z indicates the system normalized response.
Fig. 14. Comparison of the 3-5 pm system response to gaussian targets with the response to uniform rectangular targets
of corresponding diameters. The targets were computer simulated and the PSF was used for the calculations.
one-dimensional computer analysis was performed. The PSF was convolved with the normalized emissive power resulting from the simulated
temperature distribution. In Figure 14, the system
response t o targets with uniform rectangular temperature distribution is compared t o the response
to targets with gaussian temperature distribution
for several target sizes. The graph indicates the
percentage of system response t o the peak power
emitted by the target (which corresponds to the
peak temperature in the simulation). As can be
seen, the response is much lower for the gaussian
targets than for the rectangular targets of corresponding diameter. Up t o a diameter of 10 pixels,
the response to the gaussian target was about half
the response to the rectangular uniform target.
The calculated response t o gaussian targets
is presented in Figure 15 for three different peak
temperatures: 50, 100, and 150°C. For similar diameter, the higher the temperature (sharper
gaussian) the worse the response. It should be emphasized, though, that all these are responses (in
percentage) t o peak emissive power from the target, and that, due t o the non-linear relationship
between power and temperature, the percentages
do not necessarily hold for corresponding temperature rises.
atures of targets which are large and have slow
temperature variations, their accuracy is less for
small targets or rapidly changing temperatures.
Limitations are specified to a certain extent by
the manufacturers, but are often overlooked by
researchers. Particularly demanding conditions
are those occurring during tissue irradiation with
small-size laser beams for very short periods of
time. This study was directed toward evaluating
the accuracy of a camera in measuring the temperature of small and fast-changing targets like
those occurring in many medical laser applications. The study was initiated after experimental
work produced concern in our group about the
accuracy of some thermal measurements.
The detector rise time (10-90%) was found to
be 2.0 ps for the 8-12 pm camera and 5.15 p s €or
the 3-5 pm camera. The response time from 0 t o
90% of the step was 2.44 ps for the 8-12 pm system and 5.85 ps for the 3-5 pm system. It is clear
that the 3-5 pm system is more than two times
slower than the 8-12 pm system. This makes the
3-5 pm cameras particularly susceptible to errors when measuring temperatures of small targets.
The 3-5 pm thermal camera, which produces images at 30 frames per second, requires an
object 18 pixels wide (out of 256 pixels in a line) in
order
to provide the correct temperature. Thus,
DISCUSSION
the object must cover 7% of the monitor screen
Thermal cameras are being used in bio- horizontally. This corresponds t o 2.0 mm when
medical research laboratories t o estimate tissue the 3 x telescope is mounted and a close-up lens is
surface temperature during laser irradiation. Al- used t o focus the camera at 9.5 inches (24.13 cm)
though the cameras accurately measure temper- from the object. A detector response of 90% can be
zyxwv
zyxwvu
zyxwv
Limitations of a Thermal Camera
521
obtained for a target of 1.4 mm. Strictly speaking speeds. For example, some systems produce imhowever, for objects less than 2.0mm, the tem- ages at 12.5 frames per second, and even at four
perature is underestimated. This was confirmed frames per second. Since the scanning time for
by the experiment with the black body of variable one horizontal line is longer, the detector can
size heated at 100°C. The camera measured 65°C reach a higher response for small targets, and
for the 500 pm black body and 85°C for the therefore the accuracy improves. However, al1.0 mm black body. The accuracy is improved if a though slower scanning systems are spatially
close-up lens with shorter focal distance is used, more accurate, they have a poorer temporal rebecause objects appear larger in the scene and the sponse. In many cases of laser irradiation of tisdetector can reach a higher response during hor- sue, very short exposure times are used. In vessel
izontal scanning. In general, image magnification welding for example, pulses of 100 ms are typical.
with telescopes and close-up lenses improves the With a scanning rate of 30 frames per second, six
resolution and the accuracy of the measurements. fields will be scanned during this period of time,
An electro-optical zoom feature is also available while with a rate of 12.5 frames per second less
in this type of camera. This option reduces the than three fields will be scanned, and it will be
horizontal field of view by reducing the amplitude easier to miss the peak temperature that occurs
of the horizontal mirror scanning. In this way, the before the laser is turned off. Therefore, there is
image is magnified while the scene is scanned at always a trade-off between spatial accuracy and
lower speed. However, the detector instantaneous temporal resolution. A system at 4 frames per secfield of view (IFOV) does not change. The zoom ond is extremely accurate for small targets, but
only reduces errors associated with detector rise cannot be used for rapidly changing thermal
time, and it does not improve errors due to the events like those associated with laser irradiation
IFOV. The rather slow detector response from 80 of tissue.
to 100%of its final value t o a step change in temFor objects smaller than 250 pm (about 2
perature, as illustrated in Figure 6, further de- pixels), the dimensions of the detector’s IFOV are
grades the improvement in measurement accu- critical for the accuracy of the temperature mearacy for small objects which is attributed to the surements. The smaller the IFOV the better the
zoom feature.
accuracy.
In most laser applications, gaussian beam
In cases such as vessel welding with CO, laser (10.6 pm wavelength), the 8-12 pm cameras profiles (not uniform profiles) occur. Initially, becan “see” reflected laser radiation, and therefore, fore heat conduction becomes significant, these
the 3-5 pm cameras must be used t o measure gaussian beam profiles cause tissue temperature
tissue temperature. Considering the slow re- rises with the same gaussian distribution. Using
sponse of the 3-5 pm detector, tissue temperature the point spread function obtained for the 3-5 pm
measurements during vessel welding with CO, camera, the system response t o computer-simulaser are likely to be inaccurate, since the spot lated gaussian targets was found to be about half
sizes normally used vary from 200 to 500 pm. Of the response to uniform targets of corresponding
course, the effects of heat conduction must be sizes. Therefore, less accuracy is expected for
gaussian temperature profiles with small diametaken into account and will be discussed later.
ters
(< 3 mm). However, in most cases of laser
If other lasers different from the C02 laser
irradiation
of tissue, heat conduction becomes sigare used for tissue irradiation, the 8-12 pm thernificant
after
10 to 100 ms 1141. Radial heat conmal cameras are recommended, considering that
their detector system responds faster, producing duction leads to an expanded area of tissue temsharper images and providing more accurate tem- perature rise with a smooth profile. The camera
peratures for small spot sizes. Besides, this imag- then “sees” a target larger than the original laser
ing band has a more nearly linear relationship beam size, and the overall accuracy greatly imbetween radiant energy and temperature [3,131. proves. Heat conduction makes the analysis of the
The 8-12 pm camera provides correct tempera- error in the temperature measurements even
tures for targets 2900 pm when focused at 9.5 more complicated. Knowledge of the size of the
inches (using the same telescope and close-up lens thermal profile (better than the laser spot size) is
described above), with 90% detector response for of primary importance to determine the magnitude of the error and to apply correction factors.
targets of 550 pm.
In general, reports on temperature measureOther imaging systems (different from the
one used in this study) scan the scene at lower ment during CO, laser irradiation of arteries with
zyxwvuts
522
Torres et al.
small spot sizes [5-71 must be taken cautiously,
since the actual tissue temperatures may have
been much higher. That includes temperature
measurements during vessel welding. Also in relation to temperature measurement during COz
laser irradiation, Pearce et al. 131 have called attention to another important factor that affects
the measurements: the thermal gradient along
the beam. They have shown that, in the case of
tissue heated by lasers with very high absorption
coefficients like the COz laser, the temperature
gradient in the direction of propagation may have
a significant effect on the emitted radiation, and
therefore affect the thermal camera temperature
measurement. Thermal gradients within tissue to
a depth of 50-100 pm are important because the
detector receives radiation from these internal regions of the tissue. Calibration methods for the
camera are based on constant tissue temperature
conditions. If significant thermal gradients exist
in the first 100 pm below the surface, the camera
measures a temperature significantly lower than
the actual temperature at the surface of the tissue. Pearce et al. provide a correction factor t o be
multiplied by the measured emitted power. This
correction factor is (a + p.)/p, where a is the absorption coefficient of the tissue at the laser wavelength, and p is the extinction coefficient of the
imaging pass band of the camera. They have calculated a correction factor as high as 3.64 when
the 3-5 pm camera images arterial wall during
C 0 2 laser irradiation.
As a rule, the accuracy of any thermal camera must be tested prior to its use for measurements of tissue temperature during laser irradiation. Empirical correction factors can be obtained
for a specific range of temperatures. In this study,
a correction factor was derived for our 3-5 p.m
camera, on the 25-150°C temperature range,
from the experimental data plotted in Figure 12a.
By using this factor, the actual target temperature can be estimated from the measured temperature once the target size is known. It is important to point out that, due to individual
variability in detector response time and detectivity, a particular expression for an empirical correction factor must be derived for each particular
camera, and that it is unlikely that equations (5)
and (6) can be applied t o another device of the
same model.
A more precise technique, although difficult
to implement, recovers the original image by digital processing of the degraded image given by the
camera. The actual thermal image is “blurred” or
zyxw
degraded by the camera response which acts like
a “blurring filter.” If the camera response is
known, an “inverse filter” can be used to recover
the actual image from the blurred one. For this
reason, the technique is called “inverse filtering.”
To create the “inverse filter,” it is first necessary
t o find the point spread function h(x,y) of the camera. In simple terms, the theory behind this
method is the following: As discussed previously,
the resulting output image from the camera is
described as:
where f(x,y) is the unblurred original image,
h(x,y) is the system point spread function, and
g(x,y) is the blurred output image. Transforming
the above operation from the space domain to the
frequency domain we obtain:
zyxwvut
zyxwvu
G(u,v) = F(u,v) H(u,v)
(7)
where G(u,v), F(u,v) and H(u,v) are the Fourier
transforms of g(x,y), f(x,y)and h(x,y) respectively.
H(u,v) is called the “optical transfer function.”
The original image can be recovered by multiplying G(u,v) by the inverse of H(u,v) in the frequency domain, and then transforming the result
back to the space domain:
The expression l/H(u,v) = I(u,v) is called the “inverse filter.” Since I(u,v) becomes very large or is
not defined when H(u,v) approaches zero, special
manipulation of this filter is necessary.
Springer et al. [121 have developed a computer algorithm which uses this technique t o restore original thermal images of small targets
from gray scale images obtained with a thermal
camera. Targets with smooth gaussian shapes,
typical in laser applications, are handled better
by the program than targets with sharp thermal
edges. In general, peak temperatures can be recovered to a close approximation by this technique. The algorithm implemented by Springer
can be used with any image processing system
and provides the researcher with a tool to improve
the accuracy of thermal camera measurements.
zyxw
z
zyxwvutsrq
Limitations of a Thermal Camera
50°C
100°C
150°C
.
0
.
523
ablation of normal and atherosclerotic human aorta in
vitro: a first thermographic and histologic analysis. Circulation 1987; 76:1353-1563.
2. Incropera FP, DeWitt DP: “Fundamentals of Heat and
Mass Transfer,” Second Edition. New York: John Wiley &
Sons, 1985.
3. Pearce JA, Welch AJ, Motamedi M, Agah R: Thermographic measurement of tissue temperature during laser
angioplasty. In Diller KR, Roemen RB (eds): “Heat and
Mass Transfer in the Microcirculation of Thermally Significant Vessels.” Anaheim, CA: ASME, HTD, December
1986, vol 61, pp 49-54.
4. Ryu ZM: The effect of nonlinear filtering on the resolution of calibrated thermographic images. Ph.D. Dissertation, The University of Texas at Austin, May 1986.
5. Mnitentag J , Marques EF, Ribeiro MP, Braga GA, Navarro MR, Veratti AB, Armelin E, Macruz R, Jatene AD:
Thermographic study of laser on arteries. Lasers Surg
Med 1987; 7:307-329.
6. Kopchok GE, White RA, White GH, Fujitani R, Vlasak J,
Dykhovsky L, Grundfest WS: C 0 2 and argon laser vascular welding: Acute histologic and thermodynamic comparison. Lasers Surg Med 1988; 8:584-588.
7. Sorensen EMB, Thomsen S , Welch AJ, Badeau AF: Morphological and surface temperature changes in femoral
arteries following laser irradiation. Laser Surg Med
1987; 71249-257.
8. Inframetrics, Inc.: “Model 600 IR Imaging Radiometer.”
Operations manual.
9. Inframetrics, Inc.: “Model 600L IR Imaging Radiometer.”
Operator’s manual. Summer 1989.
10. Vanzetti R: “Practical Applications of Infrared Techniques.” New York: Wiley-Interscience publication, John
Wiley & Sons, 1972.
11. Rosenfeld A, Kak AC: “Digital Picture Processing,” Second Edition. 1982, Vol 1. New York: Academic Press.
12. Springer TA, Torres JH, Welch AJ, Pearce JA: Thermal
image restoration by the inverse filtering technique.
IEEE IMBS 11th Annual International Conference. November, 1989.
13. Diller KR, Pearce JA, Valvano JW, Welch AJ, Wissler
EH: Heat transfer: what’s new in bioengineering. SOMA
2(2): 14-21, July 1987.
14. van Gemert MJC, Welch AJ: Time constants in thermal
laser medicine. Lasers Surg Med 1989; 9:405-421.
zyxwvutsrqponml
zyxwvutsrqpo
,
.
.
.35 .58 .82 1.17
1.76
2.34
2.93
3.51
4.10
Diameter (mm)
Fig. 15. Calculated response to gaussian targets as a function
of target size for three different peak temperatures.
CONCLUSIONS
Infrared camera temperature measurements
of laser-irradiated tissue are subject t o several errors which lead to temperature underestimation.
For small targets, special attention must be given
to the detector instantaneous field of view as well
as to the detector rise time relative to the scanning velocity. Each thermographic imaging system must be tested to determine its accuracy and
limitations. By using empirical correction factors
or, more properly, the “inverse filtering technique,” it is possible to compensate for the errors
and to approximate the actual target temperatures.
REFERENCES
1. Welch AJ, Bradley AB, Torres JH, Motamedi M, Ghidoni
JJ, Pearce JA, Hussein H, O’Rourke RA: Laser probe
zyxw